Double-lumen tubes and auto-PEEP during one-lung ventilation
- Spaeth, J.
- Ott, M.
- Karzai, W.
- Grimm, A.
- Wirth, S.
- Schumann, S.
- Loop, T.
- Thompson, J. P.
Abstract
Background:
Double-lumen tubes (DLT) are routinely used to enable one-lung-ventilation (OLV) during thoracic anaesthesia. The flow-dependent resistance of the DLT's bronchial limb may be high as a result of its narrow inner diameter and length, and thus potentially contribute to an unintended increase in positive end-expiratory pressure (auto-PEEP). We therefore studied the impact of adult sized DLTs on the dynamic auto-PEEP during OLV.
Methods:
In this prospective clinical study, dynamic auto-PEEP was determined in 72 patients undergoing thoracic surgery, with right- and left-sided DLTs of various sizes. During OLV, air trapping was provoked by increasing inspiration to expiration ratio from 1:2 to 2:1 (five steps). Based on measured flow rate, airway pressure (Paw) and bronchial pressure (Pbronch), the pressure gradient across the DLT (ΔPDLT) and the total auto-PEEP in the respiratory system (i.e. the lungs, the DLT and the ventilator circuit) were determined. Subsequently the DLT's share in total auto-PEEP was calculated.
Results:
ΔPDLT was 2.3 (0.7) cm H2O over the entire breathing cycle. At the shortest expiratory time the mean total auto-PEEP was 2.9 (1.5) cm H2O (range 0–5.9 cm H2O). The DLT caused 27 to 31% of the total auto-PEEP. Size and side of the DLT's bronchial limb did not impact auto-PEEP significantly.
Conclusions:
Although the DLT contributes to the overall auto-PEEP, its contribution is small and independent of size and side of the DLT's bronchial limb. The choice of DLT does not influence the risk of auto-PEEP during OLV to a clinically relevant extent.
Clinical trial registration:
DRKS00005648.